A patient is breathing at a respiratory rate of 25 breaths per minute. her tidal vollume is 400ml with an I:E ratio of 1:2. She receving oxygen via a jet nebulizer set at 60% and a flow of 12L/min. Is the system adequate to meet the patient's inspiratory needs?
Ventilatory Cycle=60 seconds/ 25 BPM=2.4 seconds Inspiratory Time= 2.4 seconds/3=.8 seconds Inspiratoory Flow=0.4L/.8 seconds=.5L/sec=30L/min Entrainment Ratio at 60% = 1:1 Total Flow= 12 L/min(O2) +(12 L/min X 1(Air)=24 L/min
The patient is desaturating because she is drawing in room air (Fio2 of 0.21) at a rate of 6 L/in to meet her inspiratory demands. You must increase the flow (oxygen flowmeter) or use an alternate device (MistyOx GIN or HiFi).
Entrainment ratio indicates air to oxygen ratio, remember no matter what FiO2 you use, oxygen part is always 1. A 24% is achieved by Air:O2 of 25:1 or 35% is air to O2 ration of 4.6:1.
Above scenario, FiO2 of 60% is achieved by Air:O2=1:1, the factor of 2.
From my understanding, the ratio represents how much oxygen and air needed in order to achieve the FiO2. So, Guess what's the ratio of FiO2 100%?
The Air:O2=0:1.
BTW, I'm not familiar with the total flow calculation above,"Total Flow= 12 L/min(O2) +(12 L/min X 1(Air)=24 L/min", and I don't think that's right even though it gives the right answer of 24.
The factor is just the addition of Air/O2. Thus, 1+1=2
Another example is, FiO2 of 40%
The Air/O2 ratio is 3:1. The factor is 4.
The factor is important for finding the total flow.
Total Flow=(Flowmeter setting)x(factor). gotta memorize this formula!
So, using this formula, 12x2=24.
Anyway, The system is not helping her at all @ this point.
I'll check her O2sat or ABG, meanwhile I may wanna set higher FiO2 or CPAP maybe?!
What do you think?